Study of the relation between kVp and X-ray tube output of general X-ray machine at Hua Hin Hospital, Prachuap Khiri Khan
รหัสดีโอไอ
Creator Chawee Bunkloy
Title Study of the relation between kVp and X-ray tube output of general X-ray machine at Hua Hin Hospital, Prachuap Khiri Khan
Publisher Faculty of Associated Medical Sciences, Chiang Mai University
Publication Year 2560
Journal Title Journal of Associated Medical Sciences
Journal Vol. 50
Journal No. 3
Page no. 450-460
Keyword X-ray tube output, kVp, ESAK
URL Website https://www.tci-thaijo.org/index.php/bulletinAMS/index
Website title Journal of Associated Medical Sciences
ISSN 25396056
Abstract Introduction: X-rays are used to diagnose abnormalities in body. However, during radiographs patients will receive radiation doses. As X-ray machine has different quantities and quality of X-rays, it is therefore important for radiological technologists to set radiation doses for particular patients.Objectives: To study the X-ray tube output for general x-ray equipment and to evaluate radiation dose of patient at Hua Hin Hospital, Prachuap Khiri khan.Materials and methods: Relationship between kVp and X-ray tube output of the X-ray machine (Shimadzu HD150B-30 SN: 6114462, Philips Duo Diagnost SN: 04612 and Carestream DRX Revolution SN: 0722) was studied. Entrance surface air kerma of patients exposed from general X-ray machine was evaluated according to the guidelines of IAEA TRS No. 457.Results: Regression of output from Shimadzu HD150B-30, Philips Duo Diagnost output, and Carestream DRX Revolution was 0.003707 x kVp-0.137668, 0.003598 x kVp-0.134144, and 0.003712 x kVp -0.136235, respectively. ESAKs of Skull AP 1.37, 1.32, 1.18 and lateral 0.86, 0.83, 0.90 mGy, chest PA 0.13, 0.12, 0.06 and lateral 0.80, 0.77, 0.64 mGy, T spine AP 1.85, 1.83, 1.33 and lateral 8.80, 8.52, 7.22 mGy, abdomen AP 1.94, 1.88, 1.37 mGy, LS spine AP 2.16, 2.14, 1.55 and lateral 10.99, 10.64, 8.32 mGy, pelvis AP 1.55, 1.50, 0.96 mGy and hip AP 1.32, 1.28, 0.75 mGy were observed from Shimadzu HD150B-30, Philips Duo Diagnost and Carestream DRX Revolution, respectively.Conclusion: ESAKs of this study were lower than dose reference levels of IAEA.
Journal of Associated Medical Sciences

บรรณานุกรม

EndNote

APA

Chicago

MLA

ดิจิตอลไฟล์

Digital File
DOI Smart-Search
สวัสดีค่ะ ยินดีให้บริการสอบถาม และสืบค้นข้อมูลตัวระบุวัตถุดิจิทัล (ดีโอไอ) สำนักการวิจัยแห่งชาติ (วช.) ค่ะ