Comparative Study of HbA1c Measurement Using Immunoturbidity and High Performance Liquid Chromatography Techniques in Faculty of Medicine Vajira Hospital, Navamindradhiraj University.
รหัสดีโอไอ
Creator Lamul Wongkritwit
Title Comparative Study of HbA1c Measurement Using Immunoturbidity and High Performance Liquid Chromatography Techniques in Faculty of Medicine Vajira Hospital, Navamindradhiraj University.
Publisher Text and Journal Publication
Publication Year 2558
Journal Title Vajira Medical Journal
Journal Vol. 58
Journal No. 4
Page no. 9-16
Keyword HbA1c, immunoturbidity, HPLC
ISSN 0125-1252
Abstract Objectives: To compare HbA1c measurement methods by Immunoturbidity from Cobas Integra 800 Analyzer and High Performance Liquid Chromatography (HPLC) from ADAMS A1C HA 8180V Analyzer. Methods: This is a cross-sectional analytic study comparing two HbA1c measurement methods in the same samples. One hundred and sixty EDTA whole blood sample were determined for HbA1c by Immunoturbidity from Cobas Integra 800 Analyzer and High Performance Liquid Chromatography (HPLC) from ADAMS A1C HA8180V Analyzer. The data from two methods were separated from concentration of HbA1c: level 1 normal range HbA1c 4-6.4%, level 2 diagnose diabetes range HbA1c 6.5-8% and level 3 range HbA1c over 8 %. The data from three levels of two methods were compared and analyzed for statistical correlation. Results: Data from three levels of two methods were statistical calculated. The correlation studies of three levels showed good correlation. The correlation coefficient(r) were 0.941, 0.923 and 0.989 and linear regression equation were y=1.066x-0.451, y=0.982x+0.219 and y=1.024x- 0.017, respectively. A strong correlation was also observed. The coefficient of variation (% cv) in within run were SD=0.05, %CV=0.78 and SD=0.05, %CV=0.79 and between-run were SD= 0.08, 0.16 %CV=1.47, 1.58 and SD=0.10, 0.13 %CV=1.78, 1.58, respectively. Data from two methods showed good correlation (coefficient of variation, %CV<2%). Bland and Altman plot was used to calculate mean difference (bias) and agreement between two methods. The plot showed the presence of good agreement between two methods and three levels, data of values were lying within the mean + 1.96SD Conclusion: Statistical analysis showed that Immunoturbidity from Cobas Integra 800 Analyzer and High Performance Liquid Chromatography (HPLC) from ADAMS A1C HA8180V Analyzer were strongly correlated and congruent results in three level concentrations.
วชิรเวชสาร

บรรณานุกรม

EndNote

APA

Chicago

MLA

ดิจิตอลไฟล์

Digital File
DOI Smart-Search
สวัสดีค่ะ ยินดีให้บริการสอบถาม และสืบค้นข้อมูลตัวระบุวัตถุดิจิทัล (ดีโอไอ) สำนักการวิจัยแห่งชาติ (วช.) ค่ะ